Connecting to a Coolscope or DS-5M-L1 camera

The following describes how to directly connect a computer to either the DS-5M-L1 or Coolscope digital cameras through a network card.   

1. Use a crossover network cable.  In MS Windows open the DOS window   From the DOS  prompt type: ipconfig   This will tell you the IP Address and Subnet Mask of the PC you are connecting to. for our purposes the "connection specific DNS suffix" and the "Default Gateway" are irrelevant.

For example:

connection specific DNS suffix....micron-optics.com

IP Address....................................192.168.091.246

Subnet Mask.................................255.255.255.0

Default Gateway...........................192.168.2.1

2.Go to the Network settings menu in the camera and change the internal server Subnet Mask in the Nikon camera to the same Subnet Mask as the PC you are using.

3. Make the internal server IP Address in the Network Settings menu of the Nikon Camera the same as the IP Address for the PC except for the last digit.

IP Address....................................192.168.091.247

Subnet Mask.................................255.255.255.0

4.Save the changes in the camera.  Power the camera off and on. Type the IP Address for the camera in the address bar of Internet Explorer and hit enter.  The camera image and controls should appear in your internet browser.

 

Nikon maintains a Coolscope on continuously at their head quarters in Melville, NY.  To work with this actual system click here

 

 

Micron Optics, 240 Cedar Knolls Road , Cedar Knolls, NJ 07927

E-Mail

973-267-5799 phone

973-267-6598 fax

 

*GIF animation at top is a conoscopic image of quartz captured through an E600pol microscope using the Nikon DXM-1200 digital camera.  Courtesy of Daniel Sparling, former employee and now clergyman in training