Koehler Illumination Steps
1. Focus on the sample
with the 10x objective.
2. Close the Field Diaphragm. This is typically on the base of the microscope below the stage. Sometimes it is a knob located on the right side of the microscope (E800, E1000)
3. Focus the image of the field diaphragm to a sharp octagon using the condenser focus knob found on the left side of the microscope directly behind the condenser. The condenser is lens unit located directly below the stage.
4. Center the image of the field diaphragm using the silver knurled condenser centering knobs located at 8 o'clock and 4 o'clock directly below the condenser.
5 Open the image of the field diaphragm until it just disappears from view with the 10x objective. (Note: you will need to open it all the way if you are using objectives lower than 10x.
6. Adjust the image for optimum contrast (most pleasing to your eye) using the condenser diaphragm. The control for the condenser diaphragm is located on the front of the condenser itself. (It is located on the right side of the stand near the back on the E800M and E1000M)